Home » News » YK090 Glucagon assay performance between 100uL and 50uL specimens

Archives

News

2010.02.19
YK090 Glucagon assay performance between 100uL and 50uL specimens
YK090 Glucagon EIA Kit (Rat / Mouse / Human)

 

Introduction

For study with rat and mouse, glucagon assay with smaller volume of specimens are comfortable. YK090 Glucagon EIA can use with not only 100uL specimens but also 50uL specimens. So we compared performance of YK090 Glucagon EIA with 100uL and 50uL specimens.

 

 

Method

Glucagon assay performance between with 100uL specimens and 50uL specimens are examined. Each assay are placed along with package insert (Ⅳ. Method).

 

 

Result

Standard curve(A490)

  100uL assay 50uL assay
10000 0.009 0.044
3333 0.032 0.168
1111 0.182 0.571
370 0.646 1.228
123 1.154 1.761
41 1.398 1.990
0 1.546 2.075
Glucagon50.jpg

Glucagon value(pg/mL)

sample Human-1 Human-2 Human-3
100uL assay 602 920 254
50uL assay 716 1128 278
50/100uL assay 119% 123% 109%

 

sample Rat-1 Rat-2 Rat-3
100uL assay 1180 411 596
50uL assay 1285 265 358
50/100uL assay 109% 64% 60%

 

sample Mouse-1 Mouse-2 Mouse-3
100uL assay 638 566 298
50uL assay 593 531 252
50/100uL assay 93% 94% 85%

 

 

Discussion

Assay results with 50uL specimens are nearly same as assay results with 100uL specimens (standard procedure). But specimen volume should be fixed in a sequential study.

 

 

 